"Implementation and evaluation of the pulse field gradient technique fo" by Kevin Michael Keener
 

Implementation and evaluation of the pulse field gradient technique for the prediction of soluble solids content of fruit juices and soluble solids content and defects in apple tissue using a low frequency proton magnetic resonance sensor

Kevin Michael Keener, Purdue University

Abstract

The pulsed field gradient technique (PFG) was implemented in a 5.40 MHz $\sp1$H-MR system. A quadrupole coil was used for applying 10 to 15 G/cm gradient pulses. The self-diffusion coefficient of water, $\rm D\sb{w},$ and spin-spin relaxation time, T$\sb2,$ were measured in aqueous solutions of citric acid and/or Na-citrate buffer. Soluble solids of the solutions ranged from 0% to 16%. Juices tested and refractometer brix ranges were 0 to 11% brix in apple juice, 9 to 17% brix in red grape juice, and 21 to 34% brix in orange juice. Tissue tested was taken from Red Delicious, Golden Delicious, and Granny Smith apples. The three varieties of apples had refractometer brixes ranging from 10 to 15%. For this system, the coefficients of variation for T$\sb2$ and $\rm D\sb{w}$ measurements were 10% and 5%, respectively. Buffering and pH affected $\rm D\sb{w}$ by affecting soluble solids levels. The r$\sp2$ for linear regression of $\rm D\sb{w}$ versus refractometer brix was greater than 0.88 for all juices and aqueous solutions tested. $\rm D\sb{w}$ and T$\sb2$ were measured for bruised and unbruised tissue taken from the three apple varieties. $\rm D\sb{w}$ varied with level of both soluble and insoluble solids. Water core and internal browning did not influence $\rm D\sb{w}$ of tissue samples. However, they caused statistically significant differences ($\alpha$ = 0.05) in T$\sb2$ values. T$\sb2$ values for bruised apple tissue were 0.12 to 0.135 sec lower than T$\sb2$ values for good apple tissue. Titratable acids decreased an average of 43% in bruised compared to unbruised apple tissue. T$\sb2$ values for watercore tissue were 0.08 to 0.0935 sec lower than values for good tissue. Apple tissue with internal browning had T$\sb2$ values 0.25 seconds lower than values of good apple tissue. T$\sb2$ varied from 0.8 to 1.2 seconds in good apple tissue. The brix in a watercore apple varied with position. In the watercore tissue, the refractometer brix had an intermediate value (12.5 to 13.5%), compared to a higher brix from tissue around the water core (13 to 16%), and a lower brix from tissue farther away (10.5 to 11.5%).

Degree

Ph.D.

Advisors

Stroshine, Purdue University.

Subject Area

Agricultural engineering|Food science

Off-Campus Purdue Users:
To access this dissertation, please log in to our
proxy server
.

Share

COinS